v96 Microcode Doesn't Work on RX Devices with Five DSPs
Tech Note: TN2201
Product: TDT Drivers, RX
Version: 96
Date Added: 2022-01-07 (updated 2022-02-08)
Issue
Device microcode from the July 2021 TDT Drivers release 96 doesn't work on RX devices with 5 DSPs.
Solution
Download and install the latest TDT drivers from https://www.tdt.com/support/downloads/, then update the RX device with the following instructions:
- Turn the RX off
- Press and hold the MODE button on the front of the RX next to the power switch
- Turn the power switch on and release the MODE button 5 seconds later
- Open zBusMon
- Right click on the red
G21_1
device and select the Program option - Browse to the new RXn.dxe file
- Click on Program Device and allow the RX to update normally. This should take under 25 minutes for a 5-DSP RX device
- Retest your TDT hardware by running a Transfer Test in zBusMon
Contact TDT support support@tdt.com if you need additional assistance.